CTICU Device Study CRUISE
At a glance
Conditions
Where it's being run
1 sites across 1 statesStudy leadership
- Hernando Gomez, MD · PRINCIPAL_INVESTIGATOR · University of Pittsburgh
Who to contact
This trial hasn't published a contact. View it on ClinicalTrials.gov
Do you actually qualify for this trial?
Add a private profile and we'll compare every criterion below against your situation — and tell you which ones are met, uncertain, or excluding.
Inclusion
Exclusion
What this trial measures
- Stroke Volume Variation8 hours
The Stroke Volume Variation (SSV) will be collected continuously for 8 hours using all 3 devices (ClearSight, CareTaker and CNAP). The metric of performance will be the precision, accuracy and concordance assessed using Bland Altman analysis and Pearson's moment analysis comparing SVV from each of the non-invasive monitors with the reference standard obtained from the analysis of the arterial pressure waveform. Measurement of SVV, PPV, Eadyn and CO from non-invasive monitors will be obtained continuously, but the quantity will be recorded at different timepoints including: Baseline, every 30 min after baseline, before each bolus of crystalloid fluid or blood product (within 5 minutes of administration, after each bolus of crystalloid fluid or blood product (within 5 minutes of stopping), before each change in vasopressors or inotropes (within 5 minutes of administration), after each change in vasopressors or inotropes (within 5 minutes of stopping), before changing ventilatory mode.
- Pulse Pressure Variation (PPV)8 hours
The Pulse Pressure Variation will be collected continuously for 8 hours using all 3 devices (ClearSight, CareTaker and CNAP). The metric of performance will be the precision, accuracy and concordance assessed using Bland Altman analysis and Pearson's moment analysis comparing PPV from each of the non-invasive monitors with the reference standard obtained from the analysis of the arterial pressure waveform. This metric of performance will be reported. The most reliable device will be based on the metric of performance - that is, the most precise, accurate and concordant. Measurement of SVV, PPV, Eadyn and CO from non-invasive monitors will be obtained continuously, but the quantity will be recorded at different timepoints including: Baseline, every 30 min after baseline, before each bolus of crystalloid fluid or blood product (within 5 minutes of administration, after each bolus of crystalloid fluid or blood product (within 5 minutes of stopping), before each change in vasopressors.
- Arterial Dynamic Elastance (Eadyn)8 hours
The Arterial Dynamic Elastance (Eadyn) will be collected continuously for 8 hours using all 3 devices (ClearSight, CareTaker and CNAP). The metric of performance will be the precision, accuracy and concordance assessed using Bland Altman analysis and Pearson's moment analysis comparing Eadyn from each of the non-invasive monitors with the reference standard obtained from the analysis of the arterial pressure waveform. This metric of performance will be reported. The most reliable device will be based on the metric of performance - that is, the most precise, accurate and concordant. Measurement of SVV, PPV, Eadyn and CO from non-invasive monitors will be obtained continuously, but the quantity will be recorded at different timepoints including: Baseline, every 30 min after baseline, before each bolus of crystalloid fluid or blood product (within 5 minutes of administration, after each bolus of crystalloid fluid or blood product (within 5 minutes of stopping).